Abstract
We report on the retrofitting of a standard DP2 environmental radiation monitor replacing the photomultiplier tube with a silicon photomultiplier (SiPM). The use of a SiPM has several advantages for a hand-held radiation monitor, including convenient low-voltage operation and physical robustness. We report the detection efficiency and alpha/beta discrimination performance of the modified probe compared with an unmodified version.
Author supplied keywords
Cite
CITATION STYLE
Alharbi, M. K. M., Burgess, P., Taggart, M. P., & Sellin, P. J. (2020). Retrofitting an environmental monitor with a silicon photomultiplier sensor. Journal of Radiological Protection, 40(3), N31–N38. https://doi.org/10.1088/1361-6498/ab9312