Abstract

We report a data analysis protocol for ultrasonic velocity measurements carried out in a synchrotron x-ray facility based multi-anvil high-pressure apparatus. Synthetic ultrasonic signals for the time period between echoes from the two ends of the sample assembly are created with a few simple parameters as the echo waveform is modeled by the first echo from the anvil-assembly boundary. Each echo is modeled as a delta function with an amplitude and arrival time and convolved with this waveform. The final waveform, fit to the data, enables analysis of signals with overlapped echoes as they are common in small cell assemblies and high pressure.

Cite

CITATION STYLE

APA

Li, L., & Weidner, D. J. (2020). Full waveform analysis for high pressure ultrasonic measurement. Review of Scientific Instruments, 91(3). https://doi.org/10.1063/1.5134084