Negative reagent ions for real time detection using sift-ms

53Citations
Citations of this article
34Readers
Mendeley users who have this article in their library.

Abstract

Direct analysis techniques have greatly simplified analytical methods used to monitor analytes at trace levels in air samples. One of these methods, Selected Ion Flow Tube-Mass Spectrometry (SIFT-MS), has proven to be particularly effective because of its speed and ease of use. The range of analytes accessible using the SIFT-MS technique has been extended by this work as it introduces five new negatively charged reagent ions (O−, OH−, O2−, NO2−, and NO3−) from the same microwave powered ion source of moist air used to generate the reagent ions traditionally used (H3O+, NO+, and O2+). Results are presented using a nitrogen carrier gas showing the linearity with concentration of a number of analytes not readily accessible to positive reagent ions (CO2 from ppbv to 40,000 ppmv, sulfuryl fluoride and HCl). The range of analytes open to the SIFT-MS technique has been extended and selectivity enhanced using negative reagent ions to include CCl3NO2, SO2F2, HCN, CH3Cl, PH3, C2H4Br2, HF, HCl, SO2, SO3, and NO2.