Abstract
Objective: To compare the image quality of shoulder CT arthrography performed using 120 kVp and 140 kVp protocols. Materials and Methods: Fifty-four CT examinations were prospectively included. CT scans were performed on each patient at 120 kVp and 140 kVp; other scanning parameters were kept constant. Image qualities were qualitatively and quantitatively compared with respect to noise, contrast, and diagnostic acceptability. Diagnostic acceptabilities were graded using a one to five scale as follows: 1, suboptimal; 2, below average; 3, acceptable; 4, above average; and 5, superior. Radiation doses were also compared. Conclusion: The use of 120 kVp during shoulder CT arthrography reduces radiation dose versus 140 kVp without significant loss of image quality. Results: Contrast was better at 120 kVp, but noise was greater. No significant differences were observed between the 120 kVp and 140 kVp protocols in terms of diagnostic acceptability, signal-to-noise ratio, or contrast-to-noise ratio. Lowering tube voltage from 140 kVp to 120 kVp reduced the radiation dose by 33%.
Author supplied keywords
Cite
CITATION STYLE
Ahn, S. J., Hong, S. H., Chai, J. W., Choi, J. Y., Yoo, H. J., Kim, S. H., & Kang, H. S. (2014). Comparison of image quality of shoulder CT arthrography conducted using 120 kVp and 140 kVp protocols. Korean Journal of Radiology, 15(6), 739–745. https://doi.org/10.3348/kjr.2014.15.6.739