Simultaneous liquid-chromatographic determination of prednisone and prednisolone in plasma

13Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.
Get full text

Abstract

This high-performance liquid-chromatographic (HPLC) method for simultaneous determination of prednisone and its metabolite, prednisolone, in plasma is a modification of the method of Frey et al. (Clin Chem 1979;25:1944-7). Heparinized plasma (1.0 mL) with 0.1 mL of internal standard solution (11-deoxy-17-hydroxycorticosterone, 2 mg/L) is extracted with 7.0 mL of dichloromethane, then washed sequentially with 0.1 mol/L HCl, 0.1 mol/L NaOH, and deionized water, 2.0 mL each. The extract is evaporated and the residue reconstituted with 75 μL of mobile phase, methanol/H2O (40/60 by vol). Thirty microliters of this is injected onto a reversed-phase C6 column, which is eluted at 1.4 mL/min. Analytical recoveries of prednisone and prednisolone were 94-98% and 102-106%, respectively. Day-to-day precision (CV) was 3.8% for prednisone, 6.1% for prednisolone. We encountered no interference from the 21 other steroids and 25 drugs tested. This method is simple, accurate, and precise.

Cite

CITATION STYLE

APA

Cheng, M. H., Huang, W. Y., & Lipsey, A. I. (1988). Simultaneous liquid-chromatographic determination of prednisone and prednisolone in plasma. Clinical Chemistry, 34(9), 1897–1899. https://doi.org/10.1093/clinchem/34.9.1897