Abstract
Taper wear at the head-neck junction is a possible cause of early failure in large head metal-on-metal (LH-MoM) hip replacements. We hypothesized that: (i) taper wear may be more pronounced in certain product designs; and (ii) an increased abductor moment arm may be protective. The tapers of 104 explanted LH-MoM hip replacements revised for adverse reaction to metal debris (ARMD) from a single manufacturer were analyzed for linear and volumetric wear using a co-ordinate measuring machine. The mated stem was a shorter 12/14, threaded trunnion (n = 72) or a longer, smooth 11/13 trunnion (n = 32). The abductor moment arm was calculated from pre-revision radiographs. Independent predictors of linear and volumetric wear included taper angle, stem type, and the horizontal moment arm. Tapers mated with the threaded 12/14 trunnion had significantly higher rates of volumetric wear (0.402 mm3/yr vs. 0.123 mm3/yr [t = -2.145, p = 0.035]). There was a trend to larger abductor moment arms being protective (p = 0.055). Design variation appears to play an important role in taper-trunnion junction failure. We recommend that surgeons bear these findings in mind when considering the use of a short, threaded trunnion with a cobalt-chromium head.
Author supplied keywords
Cite
CITATION STYLE
Brock, T. M., Sidaginamale, R., Rushton, S., Nargol, A. V. F., Bowsher, J. G., Savisaar, C., … Langton, D. J. (2015). Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal-on-metal total hip arthroplasty system. Journal of Orthopaedic Research, 33(12), 1868–1874. https://doi.org/10.1002/jor.22970
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.