Abstract
The combination of the principles of two fast spectroscopic imaging (SI) methods, spectroscopic missing pulse steady-state free precession and echo planar SI (EPSI) is described as an approach toward fast 3D SI. This method, termed missing pulse steady-state free precession echo planar SI, exhibits a considerably reduced minimum total measurement time Tmin, allowing a higher temporal resolution, a larger spatial matrix size, and the use of k-space weighted averaging and phase cycling, while maintaining all advantages of the original spectroscopic missing pulse steady-state free precession sequence. The minor signal-to-noise ratio loss caused by using oscillating read gradients can be compensated by applying k-space weighted averaging. The missing pulse steady-state free precession echo planar SI sequence was implemented on a 3 T head scanner, tested on phantoms and applied to healthy volunteers. Copyright © 2011 Wiley Periodicals, Inc.
Author supplied keywords
Cite
CITATION STYLE
Dreher, W., Erhard, P., & Leibfritz, D. (2011). Fast three-dimensional proton spectroscopic imaging of the human brain at 3 T by combining spectroscopic missing pulse steady-state free precession and echo planar spectroscopic imaging. Magnetic Resonance in Medicine, 66(6), 1518–1525. https://doi.org/10.1002/mrm.22963