Colorimetric determination of ambient ozone using indigo blue droplet

21Citations
Citations of this article
51Readers
Mendeley users who have this article in their library.

Abstract

A simple and sensitive method based on a liquid droplet is described for the measurement of atmospheric ozone. A 30 μL drop of indigo blue solution is suspended in a flowing-air sampling stream. The ozone collected reacts with the indigo solution resulting in its decolorization. The colorimetric sensor is composed of two optical fibers and the source of monochromatic light was a red LED (625 nm). The calibration curve was constructed with ozone standard concentrations ranging from 37 - 123 ppbv. The detection limit achieved was 7.3 ppbv. The method considered here showed itself to be easy to apply with a fast response and a total analysis time of only 5 minutes. © 2006 Sociedade Brasileira de Química.

Cite

CITATION STYLE

APA

Felix, E. P., & Cardoso, A. A. (2006). Colorimetric determination of ambient ozone using indigo blue droplet. Journal of the Brazilian Chemical Society, 17(2), 296–301. https://doi.org/10.1590/S0103-50532006000200012