Absorption of isoflurane by Silica Gel

2Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We have studied the capacity of the drying agent silica gel (SG) to absorb isoflurane from gas samples. When dry, SG was able to absorb 31 times its own volume of isoflurane vapour, which could be recovered almost completely from the SG by displacement with water vapour. However, we were unable to demonstrate any significant absorption of isoflurane by wet SG. Care must be taken, therefore, when using SG as a drying agent in the sampling line of an analyser during research involving volatile anaesthetic agents. © 1991 British Journal of Anaesthesia.

Cite

CITATION STYLE

APA

Lumb, A. B., & Landon, M. J. (1991). Absorption of isoflurane by Silica Gel. British Journal of Anaesthesia. Oxford University Press. https://doi.org/10.1093/bja/67.1.106