Abstract
We investigated the feasibility of long-drift-time CdZnTe (CZT) gamma-ray detectors, fabricated from CZT material produced by Redlen Technologies. CZT crystals with cross-section areas of 5 × 5 mm2 and 6 × 6 mm2 and thicknesses of 20-, 30-, 40-, and 50-mm were configured as 3D position-sensitive drift detectors and were read out using a front-end ASIC. By correcting the electron charge losses caused by defects in the crystals, we demonstrated high performance for relatively thick detectors fabricated from unselected CZT material.
Cite
CITATION STYLE
Bolotnikov, A. E., Camarda, G. S., Chen, E., Cheng, S., Cui, Y., Gul, R., … James, R. B. (2016). CdZnTe position-sensitive drift detectors with thicknesses up to 5 cm. Applied Physics Letters, 108(9). https://doi.org/10.1063/1.4943161