Low flow liquid calibration setup

1Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

This article describes a primary calibration setup, and its uncertainty, for low flow liquid calibrations at Bronkhorst High-Tech. It will be used to calibrate reference flow meters from 1 to 200 g/h. By setting up an uncertainty budget for this setup, the calibration of the instruments can be compared to that of NMI's (National Metrology Institutes). The uncertainty budget consists of mass, time and mass flow uncertainties/corrections that need to be taken in to account for determining the traceable mass flow. Tests results of different flow meters/actuators measured on the setup support the calculated uncertainty. By participating in an intercomparison with NMI's the measurement and uncertainty of this setup is traceable to European NMI's.

Cite

CITATION STYLE

APA

Platenkamp, T. H., Sparreboom, W., Ratering, G. H. J. M., Katerberg, M. R., & Lötters, J. (2015). Low flow liquid calibration setup. Micromachines, 6(4), 473–486. https://doi.org/10.3390/mi6040473