Abstract
Objectives: To determine the accuracy, precision, and clinical applicability of the ICare® TONOVET Plus (TVP) in cats. Animals and Procedures: IOP readings obtained with the TVP were compared to values obtained concurrently with the original TONOVET (TV01) and Tono-Pen Vet™ (TP) in 12 normal cats (24 eyes) and 8 glaucomatous LTBP2-mutant cats (13 eyes) in vivo. Reproducibility of TVP readings was also assessed for three observers in the above cats. The anterior chambers of five different normal cat eyes were cannulated ex vivo. IOP was measured with the TVP, TV01, and TP at manometric IOPs ranging from 5 to 70 mmHg. Data were analyzed by linear regression, ANOVA and Bland–Altman plots. ANOVA was used to assess reproducibility of TVP readings obtained by different observers and an ANCOVA model controlled for variation of individual cats. p
Author supplied keywords
Cite
CITATION STYLE
Kiland, J. A., Terhaar, H. M., Walleck, H. E., Chen, N., McDaniel, K., & McLellan, G. J. (2023). Comparison of the TONOVET Plus®, TonoVet®, and Tono-Pen VetTM tonometers in normal cats and cats with glaucoma. Veterinary Ophthalmology, 26(5), 414–421. https://doi.org/10.1111/vop.13123