The effect of low-speed drilling without irrigation on heat generation: an experimental study

  • Oh J
  • Fang Y
  • Jeong S
  • et al.
N/ACitations
Citations of this article
34Readers
Mendeley users who have this article in their library.

Abstract

OBJECTIVES: In this study we evaluated heat generation during the low-speed drilling procedure without irrigation. MATERIALS AND METHODS: Ten artificial bone blocks that were similar to human D1 bone were used in this study. The baseline temperature was 37.0. We drilled into 5 artificial bone blocks 60 times at the speed of 50 rpm without irrigation. As a control group, we drilled into an additional 5 artificial bone blocks 60 times at the speed of 1,500 rpm with irrigation. The temperature changes during diameter 2 mm drilling were measured using thermocouples. RESULTS: The mean maximum temperatures during drilling were 40.9 in the test group and 39.7 in the control group. Even though a statistically significant difference existed between the two groups, the low-speed drilling did not produce overheating. CONCLUSION: These findings suggest that low-speed drilling without irrigation may not lead to overheating during drilling.

Cite

CITATION STYLE

APA

Oh, J.-H., Fang, Y., Jeong, S.-M., & Choi, B.-H. (2016). The effect of low-speed drilling without irrigation on heat generation: an experimental study. Journal of the Korean Association of Oral and Maxillofacial Surgeons, 42(1), 9. https://doi.org/10.5125/jkaoms.2016.42.1.9