Diagnostics and research analyses involving samples containing maximum-containment viruses present unique challenges, and inactivation protocols compatible with downstream testing are needed. Our aim was to identify a validated viral inactivation protocol compatible with bacterial identification by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDITOF MS). We assessed a panel of bacteria with 6 validated maximum-containment virus-inactivation protocols and report that inactivation with TRIzol or -irradiation is compatible with MALDI-TOF MS. The availability, simplicity, and rapidity of TRIzol inactivation make this method the more suitable choice.

Cite

CITATION STYLE

APA

Matson, M. J., Stock, F., Shupert, W. L., Bushmaker, T., Feldmann, F., Bishop, W. B., … Munster, V. J. (2018). Compatibility of Maximum-Containment Virus-Inactivation Protocols with Identification of Bacterial Coinfections by Matrix-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry. In Journal of Infectious Diseases (Vol. 218, pp. S297–S300). Oxford University Press. https://doi.org/10.1093/infdis/jiy292