Abstract
To research the influence of the minimum segment width (MSW) on intensity-modulated radiotherapy (IMRT) plan for left breast cancer after breast conserving surgery and provide a reference for plan optimization. Ten patients with left-sided early-stage breast cancer after breast-conserving surgery (BCS) were selected for postoperative radiotherapy. The Monaco 5.11 treatment planning system (TPS) was used, and the optimization parameters of the plan were fixed, while MSW were set to 0.5, 1, 1.5, and 2 cm, and four 5 field-IMRT plans were designed for every patient. The plan quality and execution efficiency of each plan were compared, including planning target volume (PTV) dose, organs at risk (OAR) dose, segments number (SN), monitor units (MU), delivery time (DT), and dose verification results. With the increase of the MSW, the dose indexes of the OAR were similar (P >.05). The D2%, D98%, and D50%of the PTV gradually deviated from the prescription dose, and the uniformity gradually deteriorated (P.05). The SN, MU, and DT gradually decreased (P
Author supplied keywords
Cite
CITATION STYLE
Wang, N., Chen, L., Huang, G., & Sun, H. (2023). Influence of minimum segment width on intensity-modulated radiotherapy plan for left-sided breast cancer after breast conserving surgery. Medicine (United States), 102(46), E36064. https://doi.org/10.1097/MD.0000000000036064