Clinical risk assessment of biotin interference with a high-sensitivity cardiac troponin T assay

18Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

Abstract

Objectives: Biotin >20.0 ng/mL (81.8 nmol/L) can reduce Elecsys® Troponin T Gen 5 (TnT Gen 5; Roche Diagnostics) assay recovery, potentially leading to false-negative results in patients with suspected acute myocardial infarction (AMI). We aimed to determine the prevalence of elevated biotin and AMI misclassification risk from biotin interference with the TnT Gen 5 assay. Methods: Biotin was measured using an Elecsys assay in two cohorts: (i) 797 0-h and 646 3-h samples from 850 US emergency department patients with suspected acute coronary syndrome (ACS); (ii) 2023 random samples from a US laboratory network, in which biotin distributions were extrapolated for higher values using pharmacokinetic modeling. Biotin >20.0 ng/mL (81.8 nmol/L) prevalence and biotin 99th percentile values were calculated. AMI misclassification risk due to biotin interference with the TnT Gen 5 assay was modeled using different assay cutoffs and test timepoints. Results: ACS cohort: 1/797 (0.13%) 0-h and 1/646 (0.15%) 3-h samples had biotin >20.0 ng/mL (81.8 nmol/L); 99th percentile biotin was 2.62 ng/mL (10.7 nmol/L; 0-h) and 2.38 ng/mL (9.74 nmol/L; 3-h). Using conservative assumptions, the likelihood of false-negative AMI prediction due to biotin interference was 0.026% (0-h result; 19 ng/L TnT Gen 5 assay cutoff). US laboratory cohort: 15/2023 (0.74%) samples had biotin >20.0 ng/mL (81.8 nmol/L); 99th percentile biotin was 16.6 ng/mL (68.0 nmol/L). Misclassification risk due to biotin interference (19 ng/L TnT Gen 5 assay cutoff) was 0.025% (0-h), 0.0064% (1-h), 0.00048% (3-h), and <0.00001% (6-h). Conclusions: Biotin interference has minimal impact on the TnT Gen 5 assay's clinical utility, and the likelihood of false-negative AMI prediction is extremely low.

Cite

CITATION STYLE

APA

Mumma, B., Diercks, D., Twerenbold, R., Valcour, A., Ziegler, A., Schützenmeister, A., … Tran, N. (2020). Clinical risk assessment of biotin interference with a high-sensitivity cardiac troponin T assay. Clinical Chemistry and Laboratory Medicine, 58(11), 1931–1940. https://doi.org/10.1515/cclm-2019-0962

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free