Applying three different methods of measuring CTDIfree air to the extended CTDI formalism for wide-beam scanners (IEC 60601–2–44): A comparative study

11Citations
Citations of this article
48Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: The weighted CT dose index (CTDIw) has been extended for a nominal total collimation width (nT) greater than 40 mm and relies on measurements of CTDIfree air. The purpose of this work was to compare three methods of measuring CTDIfree air and subsequent calculations of CTDIw to investigate their clinical appropriateness. Methods: The CTDIfree air, for multiple nTs up to 160 mm, was calculated from (1) high-resolution air kerma profiles from a step-and-shoot translation of a liquid ionization chamber (LIC) (considered to be a dosimetric reference), (2) pencil ionization chamber (PIC) measurements at multiple contiguous positions, and (3) air kerma profiles obtained through the continuous translation of a solid-state detector. The resulting CTDIfree air was used to calculate the CTDIw, per the extended formalism, and compared. Results: The LIC indicated that a 40 mm nT should not be excluded from the extension of the CTDIw formalism. The solid-state detector differed by as much as 8% compared to the LIC. The PIC was the most straightforward method and gave equivalent results to the LIC. Conclusions: The CTDIw calculated with the latest CTDI formalism will differ most for 160 mm nTs (e.g., whole-organ perfusion or coronary CT angiography) compared to the previous CTDI formalism. Inaccuracies in the measurement of CTDIfree air will subsequently manifest themselves as erroneous calculations of the CTDIw, for nTs greater than 40 mm, with the latest CTDI formalism. The PIC was found to be the most clinically feasible method and was validated against the LIC.

Cite

CITATION STYLE

APA

Bujila, R., Kull, L., Danielsson, M., & Andersson, J. (2018). Applying three different methods of measuring CTDIfree air to the extended CTDI formalism for wide-beam scanners (IEC 60601–2–44): A comparative study. Journal of Applied Clinical Medical Physics, 19(4), 281–289. https://doi.org/10.1002/acm2.12363

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free