Evaluation of light intensity output of QTH and LED curing devices in various governmental health institutions

37Citations
Citations of this article
51Readers
Mendeley users who have this article in their library.

Abstract

The purpose of this study was to evaluate the light intensity output of quartz-tungsten-halogen (QTH) and light emitting diode (LED) curing devices located at governmental health institutions in Riyadh, Saudi Arabia. Eight governmental institutions were involved in the study. The total number of evaluated curing devices was 210 (120 were QTH and 90 were LED). The reading of the light intensity output for each curing unit was achieved using a digital spectrometer; (Model USB4000 Spectrometer, Ocean Optics Inc, Dunedin, FL, USA). The reading procedure was performed by a single investigator; any recording of light intensity below 300 mW/cm2 was considered unsatisfactory. The result found that the recorded mean values of light intensity output for QTH and LED devices were 2±0 mW/cm2 and 598 mW/ cm2, respectively. The percentage of QTH devices and LED devices considered unsatisfactory was ±7.5% and 15.±%, respectively. Overall, the regular assessment of light curing devices using light meters is recommended to assure adequate output for clinical use. © Operative Dentistry, 2011.

Cite

CITATION STYLE

APA

Al Shaafi, M. M., Maawadh, A. M., & Al Qahtani, M. Q. (2011). Evaluation of light intensity output of QTH and LED curing devices in various governmental health institutions. Operative Dentistry, 36(4), 356–361. https://doi.org/10.2341/10-247-O

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free