Abstract

The Reference Intensity Ratio (RIR) is a general, instrument-independent constant for use in quantitative phase analysis by the X-ray powder diffraction internal standard method. When the reference standard is corundum, RIR is known as I/Ic; These constants are collected in the Powder Diffraction File (1987), can be calculated, and can be measured. Recommended methods for accurate measurement of RIR constants are presented, and methods of using these constants for quantitative analysis are discussed. The numerous, complex constants in Copeland and Bragg's method introduced to account for superimposed lines can be simply expressed in terms of RIR constants and relative intensities. This formalism also permits introduction of constraints and supplemental equations based on elemental analysis. © 1988, Cambridge University Press. All rights reserved.

Cite

CITATION STYLE

APA

Hubbard, C. R., & Snyder, R. L. (1988). RIR — Measurement and Use in Quantitative XRD. Powder Diffraction, 3(2), 74–77. https://doi.org/10.1017/S0885715600013257