Rate Dependency During Relaxation of Superelastic Orthodontic NiTi Alloys After Hydrogen Charging

12Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The relaxation behavior under tensile loading of a superelastic NiTi alloy was investigated after hydrogen charging with respect to aging from one to 77 days in air at room temperature. The specimens were immersed for 3 h in a 0.9 % NaCl aqueous solution and then relaxed with an imposed strain of 4.8 %—which results in half of the martensite transformation—for different strain rates of 10−4, 10−3, and 5 × 10−3 s−1. For the non-charged specimens, the relaxed stress at the beginning exhibited a temporary dependence on the strain rates and then reached the same equilibrium stress after 2.5 h. After hydrogen charging, this equilibrium stress did not vary for the as-charged specimen. Nevertheless, the greater the aging period is the greater the equilibrium stress is. This behavior can be attributed to the diffusion of hydrogen into the entire specimen, which hinders the relaxation mechanism of the martensite bands.

Cite

CITATION STYLE

APA

Elkhal Letaief, W., Hassine, T., & Gamaoun, F. (2016). Rate Dependency During Relaxation of Superelastic Orthodontic NiTi Alloys After Hydrogen Charging. Shape Memory and Superelasticity, 2(1), 121–127. https://doi.org/10.1007/s40830-016-0051-6

Register to see more suggestions