Abstract
Membranes of 4-tert-butylcalix(8)arene-octaacetic acid octaethyl ester (I) as an electroactive material, sodium tetraphenyl borate (NaTPB) as an arrion excluder, and tri-n-butyl phosphate (TBP) as a solvent mediator in poly(vinyl chloride) (PVC) matrix have been tried for a strontium-selective sensor. The best performance was exhibited by the membrane having a composition 5:100:150:2 (I: PVC: TBP: NaTPB (w/w)). This sensor exhibits a good potentiometric response to Sr2+ over a wide concentration range (3.2 × 10-5 -1.0 × 10-1 M) with a Nernstian slope (30 mV/ decade). The response time of the sensor is 10 s and it has been used for a period of four months without any drift in potentials. The selectivity coefficient values are in the order of 0.01 for mono-, bi-, and trivalent cations which indicate a good selectivity for Sr2+ over a large number of cations. The useful pH range for the sensor was found to be 3-10 and it works well in mixtures with non-aqueous content up to 25 % (v/v). The sensor has been used as an indicator electrode in the potentiometric titration of Sr2+ against EDTA. ©2004 by MDPI.
Author supplied keywords
Cite
CITATION STYLE
Jain, A. K., Gupta, V. K., & Raisoni, J. R. (2004). Strontium (II)-selective potentiometric sensor based on ester derivative of 4-tert-butylcalix(8)arene in PVC matrix. Sensors, 4(8), 115–124. https://doi.org/10.3390/s40800115