Calibration of the gain and digital conversion factor of an EMCCD is necessary for accurate photon counting. We present a new method to quickly calibrate multiple gain settings of an EMCCD camera. Acquiring gain-series calibration data and analyzing the resulting images with the EMCCD noise model more accurately estimates the gain response of the camera. Furthermore, we develop a method to compare the results from different calibration approaches. Gain-series calibration outperforms all other methods in this self-consistency test.
CITATION STYLE
Ryan, D. P., Dunlap, M. K., Gelfand, M. P., Werner, J. H., Van Orden, A. K., & Goodwin, P. M. (2021). A gain series method for accurate EMCCD calibration. Scientific Reports, 11(1). https://doi.org/10.1038/s41598-021-97759-6