Abstract
Sevoflurane is degraded by soda lime to a vinyl ether commonly referred to as compound A. We measured the concentration of compound A in the circle breathing system of 31 patients receiving seroflurane anasthesia. Inspiratory and expiratory gas samples were analysed using gas chromatography and flame ionisation detection. The end-tidal seroflurane concentration and soda lime temperature were recorded. The peak compound A concentration ranged between 10 to 32 ppm in the inspiratory limb and 7 to 26ppm in the expiratory limb. There was a positive correlation between the peak compound A concentration and the end-tidal sevoflurane concentration (r2 = 0.545, p < 0.0001) and the soda lime temperature (r2 = 0.301, p = 0.0014). We conclude that the end-tidal concentration of sevoflurane and the temperature of the soda lime are important variables in determining concentration of compound A in a circle system.
Author supplied keywords
Cite
CITATION STYLE
Munday, I. T., Ward, P. M., Foden, N. D., Jones, R. M., Van Pelt, F. N. A. M., & Kenna, J. G. (1996). Sevoflurane degradation by soda lime in a circle breathing system. Anaesthesia, 51(7), 622–626. https://doi.org/10.1111/j.1365-2044.1996.tb07842.x