Abstract
A flow injection preconcentration system for the flame atomic absorption spectrometric determination of hexavalent chromium has been developed. The method employs on-line preconcentration of Cr(VI) on a minicolumn packed with Cr(VI)-imprinted poly(4-vinyl pyridineco-2-hydroxyethyl methacrylate) placed into a flow injection system. Hexava-lent chromium was eluted with a small volume of diluted hydrochloric acid into the nebulizer-burner system of a flame atomic absorption spectrometer. An enrichment factor of 550 and a 3σ detection limit of 0.04 μg·L-1 along a sampling frequency of 4 h-1 at a sample flow rate of 3.5 mL·min-1. The relative standard deviation is 2.9% for 1 μg·L-1 Cr(VI) (n = 11). The flow injection system proposed has the advantage of being simpler because the use of expensive and sophisticated instruments is avoided. Ease of use, continuous process and selectivity make this method suitable for Cr(VI) determination in different environmental samples such as sea and river waters, soils and sediments.
Cite
CITATION STYLE
Yebra-Biurrun, M. C., Castro-Romero, J. M., & Carro-Mariño, N. (2012). Flow-Injection Flame Atomic Absorption Determination of Hexavalent Chromium with On-Line Preconcentration on an Anion Imprinted Polymer. American Journal of Analytical Chemistry, 03(11), 755–760. https://doi.org/10.4236/ajac.2012.311100