Abstract
Haemodynamic instability is a recognised phenomenon in critically ill patients undergoing hyperbaric therapy. Instability may result from the effects of ambient pressure on the cardiovascular system, devices involved in infusion of drugs and fluids, or a combination of the two. The effect of hyperbaric pressure on air-containing spaces in syringes has not been previously measured. We connected 60-ml syringes (Terumo Corporation, Tokyo, Japan) containing coloured water to low volume extensions via three-way taps. We examined the effect of pressurisation to 2.4 and 2.8 atmospheres absolute (ATA) on the syringes by measuring the displacement of the coloured water in the low volume extension set. There was compression of air spaces within the syringe causing retrograde flow of fluid within the low volume extension set. The mean (95% CI) change in volume was 154 (141-168) μl at 2.4 ATA, and 197 (183-212) μl at 2.8 ATA (both p < 0.0001). We conclude that hyperbaric exposure may cause clinically significant changes in syringe function at infusion rates < 100 ml. h-1. © 2007 The Authors Journal compilation 2007 The Association of Anaesthetists of Great Britain and Ireland.
Cite
CITATION STYLE
Hopson, A. S. M., & Greenstein, A. (2007). Intravenous infusions in hyperbaric chambers: Effect of compression on syringe function. Anaesthesia, 62(6), 602–604. https://doi.org/10.1111/j.1365-2044.2007.05027.x