Noninvasive uterine electromyography for prediction of preterm delivery

164Citations
Citations of this article
141Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Objective Power spectrum (PS) of uterine electromyography (EMG) can identify true labor. EMG propagation velocity (PV) to diagnose labor has not been reported. The objective was to compare uterine EMG against current methods to predict preterm delivery. Study Design EMG was recorded in 116 patients (preterm labor, n = 20; preterm nonlabor, n = 68; term labor, n = 22; term nonlabor, n = 6). A Student t test was used to compare EMG values for labor vs nonlabor (P < .05, significant). Predictive values of EMG, Bishop score, contractions on tocogram, and transvaginal cervical length were calculated using receiver-operator characteristics analysis. Results PV was higher in preterm and term labor compared with nonlabor (P < .001). Combined PV and PS peak frequency predicted preterm delivery within 7 days with area under the curve (AUC) of 0.96. Bishop score, contractions, and cervical length had an AUC of 0.72, 0.67, and 0.54. Conclusion Uterine EMG PV and PS peak frequency more accurately identify true preterm labor than clinical methods. © 2011 Mosby, Inc. All rights reserved.

Cite

CITATION STYLE

APA

Lucovnik, M., Maner, W. L., Chambliss, L. R., Blumrick, R., Balducci, J., Novak-Antolic, Z., & Garfield, R. E. (2011). Noninvasive uterine electromyography for prediction of preterm delivery. In American Journal of Obstetrics and Gynecology (Vol. 204, pp. 228.e1-228.e10). Mosby Inc. https://doi.org/10.1016/j.ajog.2010.09.024

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free