Drug output from nebulizers is dependent on the method of measurement

29Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The objective of this study was to determine whether current regulatory methods for assessing the output of nebulizers are appropriate for the delivery of nebulized steroid suspensions to patients. We studied a conventional jet nebulizer (the Intersurgical Cirrus), an open-vent nebulizer (the Medicaid Sidestream) and a breath-enhanced nebulizer (the Pari LC Plus), using a constant sampling flow or a sinusoidal pump to represent the breathing pattern of children from 6 months to adulthood. Recovery of budesonide released from the nebulizers onto filters was reduced when using breathing simulation compared with constant flow, and this reduction was greatest for the conventional nebulizer (Cirrus, 103 μg with constant flow to 4.4 μg with a 50 mL tidal volume; Pari, 176 μg to 25 μg). The open- vent nebulizer deposited very little budesonide on the filter at lower tidal volumes (4.5 μg with a 50 mL tidal volume), possibly because the enhanced flow of aerosol laden air was greater than the inspiratory flow from the breathing simulator. The output of the LC plus was reduced at high flow, from 176 μg at 20 L·min-1 to 93 μg at 60 L·min-1. Overall, the measured output varied by up to 700%, depending on the method used. These results suggest that breathing patterns dramatically alter the measured output of different nebulizers and that breathing simulation should he included as part of their assessment.

Cite

CITATION STYLE

APA

Barry, P. W., & O’Callaghan, C. (1998). Drug output from nebulizers is dependent on the method of measurement. European Respiratory Journal, 12(2), 463–466. https://doi.org/10.1183/09031936.98.12020463

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free