Enhancement effects and relaxivities of gadolinium-DTPA at 1.5 versus 3 Tesla: A phantom study

83Citations
Citations of this article
83Readers
Mendeley users who have this article in their library.

Abstract

Purpose: To investigate the difference in enhancement effects and relaxivities of the gadolinium chelate at 1.5 and 3 Tesla (T) and to elucidate the contribution of the high magnetic field to contrast enhancement in spin-echo (SE) and gradient-echo (GRE) images. Methods: Phantoms containing water with or without gadopentetate dimeglumine (Gd-DTPA) at different concentrations were scanned using 1.5T and 3T MRI scanners of the same manufacturer and under the same temperature conditions and scanning parameters. Relaxivities of gadolinium, R1 and R2, were estimated from serial T1 and T2 values of the phantoms using linear regression. Contrast enhancement ratios in SE and GRE T1-weighted images were compared at 1.5 and 3T. Results: The R1 and R2 of Gd-DTPA at 1.5 and 3T were 4.79 and 5.14, and 4.50 and 5.09, respectively. Although the relaxivities at 3T were slightly lower than those at 1.5T, the contrast enhancement ratio improved in both SE and GRE images as a result of T1 prolongation of the water at 3T. Conclusion: The decrease in relaxivities of the Gd-DTPA at 3T appears to be so small that T1 prolongation of the water improves contrast enhancement, suggesting a potential clinical advantage in administration of Gd-DTPA at high field strength.

Cite

CITATION STYLE

APA

Sasaki, M., Shibata, E., Kanbara, Y., & Ehara, S. (2005). Enhancement effects and relaxivities of gadolinium-DTPA at 1.5 versus 3 Tesla: A phantom study. Magnetic Resonance in Medical Sciences, 4(3), 145–149. https://doi.org/10.2463/mrms.4.145

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free