Abstract
The "transXend" detector measures X-rays as electric currents and provides the energy distribution of the measured X-rays after analysis. Capabilities of material distinction, effective atomic number measurement, and low-dose exposure computed tomography (CT) with high K-edge contrast agent from the use of the transXend detector in energy-resolved CT have been demonstrated via the first-generation CT measurements. For application of the principle of the transXend detector to the third-generation CT for human subjects in future work, a method for fabrication of a two-dimensional transXend detector is proposed and demonstrated using a commercial two-dimensional detector and two kinds of strip absorbers. The energy-resolved CT is performed by placing a proposed absorber system in front of a two-dimensional detector, which is used for conventional current measurement CT.
Author supplied keywords
Cite
CITATION STYLE
Kanno, I., Yamashita, Y., Kanai, E., Ogawa, T., & Shinsho, K. (2016). Two-dimensional “transXend” detector for third-generation energy-resolved computed tomography. Journal of Nuclear Science and Technology, 53(2), 258–262. https://doi.org/10.1080/00223131.2015.1037810