Abstract
Objective: To evaluate the clinical performance of the OPPO Watch (OW) Sleep Analyzer (OWSA) on OSA screening with polysomnography reference. Methods: We recruited 350 participants using OWSA and PSG simultaneously in a sleep laboratory. The respiratory event index (REI) derived from OWSA and the apnea-hypopnea index (AHI) provided by PSG were compared. SHapley Additive exPlanation (SHAP) values were calculated to explain the model of OWSA. Results: The OWSA-REI (26.5±18.5 events/h) correlated well with PSG-AHI (33.2±25.7 events/h; r = 0.91, p < 0.001), with an intraclass correlation coefficient (ICC) of 0.83. Using a threshold of AHI ≥15 events/h, the sensitivity, specificity, accuracy, and area under the curve (AUC) were 86.1%, 86.7%, 86.3%, and 0.94, respectively. Bland-Altman analysis showed that OWSA-REI and PSG-AHI were in good agreement (Mean Difference: −6.7, 95% CI:16.0 to −29.3 events/h). In addition, the effectiveness of the models in OWSA were also explained by visualizing SHAP values. Conclusion: The OWSA demonstrated a reasonable performance for OSA screening in the clinical setting. In light of this, it is possible for smartwatches to become a complementary tool to PSG, which is particularly useful for larger-scale preliminary screenings.
Author supplied keywords
Cite
CITATION STYLE
Zhou, G., Zhao, W., Zhang, Y., Zhou, W., Yan, H., Wei, Y., … Cheng, H. (2024). Comparison of OPPO Watch Sleep Analyzer and Polysomnography for Obstructive Sleep Apnea Screening. Nature and Science of Sleep, 16, 125–141. https://doi.org/10.2147/NSS.S438065