Harmonization of standard uptake values across different positron emission tomography/computed tomography systems and different reconstruction algorithms: validation in oncology patients

8Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: EQ.PET is a software package that overcomes the reconstruction-dependent variation of standard uptake values (SUV). In this study, we validated the use of EQ.PET for harmonizing SUVs between different positron emission tomography/computed tomography (PET/CT) systems and reconstruction algorithms. Methods: In this retrospective study, 49 patients with various cancers were scanned on a Biograph mCT (mCT) or Gemini TF 16 (Gemini) after [18F]FDG injections. Three groups of patient data were collected: Group 1, patients scanned on mCT or Gemini with data reconstructed using two parameters; Group 2, patients scanned twice on different PET scanners (interval between two scans, 68.9 ± 41.4 days); and Group 3, patients scanned twice using mCT with data reconstructed using different algorithms (interval between two scans, 109.5 ± 60.6 days). The SUVs of the lesions and background were measured, and the tumor-to-background ratios (TBRs) were calculated. In addition, the consistency between the two reconstruction algorithms and confounding factors were evaluated. Results: In Group 1, the consistency of SUV and TBR between different reconstruction algorithms improved when the EQ.PET filter was applied. In Group 2, by comparing ΔSUV, ΔSUV%, ΔTBR, and ΔTBR% with and without the EQ.PET, the results showed significant differences (P < 0.05). In Group 3, Bland–Altman analysis of ΔSUV with EQ.PET showed an improved consistency relative to that without EQ.PET. Conclusions: EQ.PET is an efficient tool to harmonize SUVs and TBRs across different reconstruction algorithms. Patients could benefit from the harmonized SUV, ΔSUV, and ΔSUV% for therapy responses and follow-up evaluations.

Author supplied keywords

Cite

CITATION STYLE

APA

Song, Y., Meng, X., Cao, Z., Zhao, W., Zhang, Y., Guo, R., … Li, N. (2023). Harmonization of standard uptake values across different positron emission tomography/computed tomography systems and different reconstruction algorithms: validation in oncology patients. EJNMMI Physics, 10(1). https://doi.org/10.1186/s40658-023-00540-z

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free