Use of diverging apertures to minimize the edge scatter in passive scattering proton therapy

6Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The purpose of this study was to evaluate the use of diverging-cut aperture to minimize collimator contamination in proton therapy. Two sets of apertures with nondivergent and divergent edge were fabricated to produce a 10cm × 10cm field at the radiation isocenter of a single-room proton therapy unit. Transverse profiles were acquired in a scanning water tank with both aperture sets. Up to 9.5% extra dose was observed from aperture scattering near the field edges with the nondivergent aperture set at 2cm above the water surface and remained 3.0% at depth of 10cm. For the divergent set, the contamination was reduced to less than 3.5% and 1.3%, respectively. Our study demonstrated that scattering from apertures contaminated the dose distribution near the field edge at shallow depth. A diverging-cut aperture was capable of reducing the contamination and is recommended for use in passive scattering proton therapy, especially when critical organs are lateral and proximal to the target at shallow depth.

Cite

CITATION STYLE

APA

Zhao, T., Cai, B., Sun, B., Grantham, K., Mutic, S., & Klein, E. (2015). Use of diverging apertures to minimize the edge scatter in passive scattering proton therapy. Journal of Applied Clinical Medical Physics, 16(5), 367–372. https://doi.org/10.1120/jacmp.v16i5.5675

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free