A novel method for conformity assessment testing of therapeutic ultrasounds for post-market surveillance purposes

1Citations
Citations of this article
76Readers
Mendeley users who have this article in their library.

Abstract

BACKGROUND: Therapeutic ultrasounds are medical devices used for treatment of conditions such muscle spasms, joint contractures or general muscle pain. Their function relies in the delivery of ultrasonic pulses that generate heat in tissue thus relieving the symptoms of aforementioned conditions. Accuracy of the delivered pulses directly affects the quality and effectiveness of the treatment and has to be ensured throughout the utilization of the therapeutic ultrasound in practice. The new Medical Device Regulation (MDR) defines medical device post-market surveillance (PMS) as performed by independent, third-party, notified bodies more strategically in hope to improve traceability of device performance. However, there is still an apparent gap in terms of standardised conformity assessment testing methods. OBJECTIVE: This paper proposes a novel method for conformity assessment testing of therapeutic ultrasounds for post-market surveillance purposes. METHOD: The method was developed based on metrology characteristics of therapeutic ultrasounds and includes visual, electrical safety and performance inspections of therapeutic ultrasounds to ensure that both safety and treatment reliability are achieved. RESULTS: The developed method was validated between 2018 and 2021 in healthcare institutions of all levels. The results obtained during validation suggest that conformity assessment testing of therapeutic ultrasounds as a method used during PMS contributes to significant improvement in devices' accuracy and reliability. CONCLUSION: A standardized approach in conformity assessment testing of therapeutic ultrasounds during PMS, besides increasing reliability of the devices, is the first step in digital transformation of management of these devices in healthcare institutions opening possibility for use of artificial intelligence.

Cite

CITATION STYLE

APA

Badnjevic, A., Deumic, A., Ademovic, A., & Pokvic, L. G. (2023). A novel method for conformity assessment testing of therapeutic ultrasounds for post-market surveillance purposes. Technology and Health Care, 31(1), 339–346. https://doi.org/10.3233/THC-229009

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free