Abstract
Kilovoltage cone beam computed tomography (CBCT) imaging may be useful in verifying patient position in stereotactic radiosurgery (SRS). To evaluate its efficacy, we investigated isocenter differences in the radiation beam and CBCT with respect to the achievable setup of a conventional frame-based SRS system. A verification phantom constructed from two plastic boards and Gafchromic-EBT film (4 × 4 cm2) pricked with a pin, was scanned by simulation CT. An isocenter at the tip of pin was planned in the treatment planning system and positioned using stereotactic coordinates. Star-shot irradiation was performed to evaluatethe difference between the radiation isocenter and the target (pinhole). CBCT rotation of 200° with a micro multileaf collimator (m3) was performed and measured the isocenter difference between CBCT and the target (tip of pin) by comparing relative coordinates. Data acquisitionwas performed 13 times on different days and differences were analyzed by calculating mean andstandard deviation. The mean difference between the radiation beam and the target (pinhole) and between radiation beam and CBCT isocenter, were 0.6 ± 0.2 mm and 0.8 ± 0.1 mm,respectively. The setup accuracy of conventional stereotactic coordinates and the isocenter accuracy of CBCT complied with AAPM Report No. 54.
Author supplied keywords
Cite
CITATION STYLE
Fukuda, A. (2010). Pretreatment setup verification by cone beam CT in stereotactic radiosurgery: Phantom study. Journal of Applied Clinical Medical Physics, 11(4), 122–129. https://doi.org/10.1120/jacmp.v11i4.3162
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.