Dosimetric characterization of a synthetic single crystal diamond detector in clinical radiation therapy small photon beams

105Citations
Citations of this article
110Readers
Mendeley users who have this article in their library.

Abstract

Purpose: To determine the potentialities of synthetic single crystal diamond Schottky diodes for accurate dose measurements in radiation therapy small photon beams. Methods: The dosimetric properties of a diamond-based detector were assessed by comparison with a reference microionization chamber. The diamond device was operated at zero bias voltage under irradiation with high-energy radiotherapic photon beams. The stability of the detector response and its dose and dose rate dependence were measured. Different square field sizes ranging from 1 × 1 cm2 to 10 × 10 cm2 were used during comparative dose distribution measurements by means of percentage depth dose curves (PDDs), lateral beam profiles, and output factors. The angular and temperature dependence of the diamond detector response were also studied. Results: The detector response shows a deviation from linearity of less than ±0.5 in the 0.01-7 Gy range and dose rate dependence below ±0.5 in the 1-6 Gymin range. PDDs and output factors are in good agreement with those measured by the reference ionization chamber within 1. No angular dependence is observed by rotating the detector along its axis, while ∼3.5 maximum difference is measured by varying the radiation incidence angle in the polar direction. The temperature dependence was investigated as well and a ±0.2 variation of the detector response is found in the 18-40 °C range. Conclusions: The obtained results indicate the investigated synthetic diamond-based detector as a candidate for small field clinical radiation dosimetry in advanced radiation therapy techniques. © 2012 American Association of Physicists in Medicine.

Cite

CITATION STYLE

APA

Ciancaglioni, I., Marinelli, M., Milani, E., Prestopino, G., Verona, C., Verona-Rinati, G., … De Notaristefani, F. (2012). Dosimetric characterization of a synthetic single crystal diamond detector in clinical radiation therapy small photon beams. Medical Physics, 39(7), 4493–4501. https://doi.org/10.1118/1.4729739

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free