Abstract
The National Ignition Facility's full-aperture backscatter station (FABS) is described. The FABS uses five independent diagnostics on each of the four laser beams in the initial National Ignition Facility quad to measure the energy, power, spectrum, and near-field amplitude modulations of the stimulated Brillouin and stimulated Raman backscattered light. In initial tests CO 2 and C 5H 12 gas-filled targets were used to create various laser-plasma interaction conditions which have shown the capability of producing ignition size laser plasmas with reflectivites on the order of 10%. Results are presented for tests in which 16 kJ on target produced between 0.3 and 2.5 kJ of backscattered light. © 2004 American Institute of Physics.
Cite
CITATION STYLE
Froula, D. H., Bower, D., Chrisp, M., Grace, S., Kamperschroer, J. H., Kelleher, T. M., … Glenzer, S. H. (2004). Full-aperture backscatter measurements on the National Ignition Facility. In Review of Scientific Instruments (Vol. 75, pp. 4168–4170). https://doi.org/10.1063/1.1789592