Abstract
We report on the design, performance assessment, and first in vivo measurement of a Time-Resolved Diffuse Optical system for broadband (600-1350 nm) nm measurement of absorption and scattering spectra of biological tissues for non-invasive clinical diagnostics. Two strategies to reduce drift and enhance responsivity are adopted. The system was enrolled in a first in vivo test phase on healthy volunteers, carrying out non-invasive, in vivo quantification of key tissue constituents (oxy- and deoxy-hemoglobin, water, lipids, collagen) and tissue micro-structure (scatterer size and density).
Author supplied keywords
Cite
CITATION STYLE
Sekar, S. K. V., Farina, A., Martinenghi, E., Mora, A. D., Taroni, P., Pifferi, A., … Urbano-Ispizua, A. (2015). Broadband time-resolved diffuse optical spectrometer for clinical diagnostics: Characterization and in-vivo measurements in the 600-1350 nm spectral range. In Optics InfoBase Conference Papers. Optica Publishing Group (formerly OSA). https://doi.org/10.1117/12.2183835