Abstract
Purpose: Accurate and artifact-free T1ρ quantification is still a major challenge due to a susceptibility of the spin-locking module to B0 and/or B1 field inhomogeneities. In this study, we present a novel spin-lock preparation module (B-SL) that enables an almost full compensation of both types of inhomogeneities. Methods: The new B-SL module contains a second 180° refocusing pulse to compensate each pulse in the preparation block by a corresponding pulse with opposite phase. For evaluation and validation of B-SL, extensive simulations as well as phantom measurements were performed. Furthermore, the new module was compared to three common established compensation methods. Results: Both simulations and measurements demonstrate a much lower susceptibility to artifacts for the B-SL module, therefore providing an improved accuracy in T1ρ quantification. In the presence of field inhomogeneities, measurements revealed an increased banding compensation by 79% compared with the frequently used composite module. The goodness of the mono-exponential T1ρ fitting procedure was improved by 58%. Conclusion: The B-SL preparation enables the generation of accurate relaxation maps with significantly reduced artifacts, even in the case of large field imperfections. Therefore, the B-SL module is suggested to be highly beneficial for in vivo T1ρ quantification.
Author supplied keywords
Cite
CITATION STYLE
Gram, M., Seethaler, M., Gensler, D., Oberberger, J., Jakob, P. M., & Nordbeck, P. (2021). Balanced spin-lock preparation for B1-insensitive and B0-insensitive quantification of the rotating frame relaxation time T1ρ. Magnetic Resonance in Medicine, 85(5), 2771–2780. https://doi.org/10.1002/mrm.28585
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.