Chiroptical measurement of chiral aggregates at liquid-liquid interface in centrifugal liquid membrane cell by mueller matrix and conventional circular dichroism methods

15Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

Abstract

The centrifugal liquid membrane (CLM) cell has been utilized for chiroptical studies of liquid-liquid interfaces with a conventional circular dichroism (CD) spectropolarimeter. These studies required the characterization of optical properties of the rotating cylindrical CLM glass cell, which was used under the high speed rotation. In the present study, we have measured the circular and linear dichroism (CD and LD) spectra and the circular and linear birefringence (CB and LB) spectra of the CLM cell itself as well as those of porphyrine aggregates formed at the liquid-liquid interface in the CLM cell, applying Mueller matrix measurement method. From the results, it was confirmed that the CLM-CD spectra of the interfacial porphyrin aggregates observed by a conventional CD spectropolarimeter should be correct irrespective of LD and LB signals in the CLM cell. © 2011 by the authors.

Cite

CITATION STYLE

APA

Takechi, H., Arteaga, O., Ribo, J. M., & Watarai, H. (2011). Chiroptical measurement of chiral aggregates at liquid-liquid interface in centrifugal liquid membrane cell by mueller matrix and conventional circular dichroism methods. Molecules, 16(5), 3636–3647. https://doi.org/10.3390/molecules16053636