Interplay effect of target motion and pencil-beam scanning in proton therapy for pediatric patients

23Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: To investigate the effect of interplay between spot-scanning proton beams and respiration-induced tumor motion on internal target volume coverage for pediatric patients. Materials and Methods: Photon treatments for 10 children with representative tumor motions (1-13 mm superior-inferior) were replanned to simulate single-field uniform dose- optimized proton therapy. Static plans were designed by using average computed tomography (CT) data sets created from 4D CT data to obtain nominal dose distributions. The motion interplay effect was simulated by assigning each spot in the static plan delivery sequence to 1 of 10 respiratory-phase CTs, using the actual patient breathing trace and specifications of a synchrotron-based proton system. Dose distributions for individual phases were deformed onto the space of the average CT and summed to produce the accumulated dose distribution, whose dose-volume histogram was compared with the one from the static plan. Results: Tumor motion had minimal impact on the internal target volume hot spot (D2), which deviated by ,3% from the nominal values of the static plans. The cold spot (D98) was also minimally affected, except in 2 patients with diaphragmatic tumor motion exceeding 10mm. The impact on tumor coverage wasmore pronounced with respect to the V99 rather than the V95. Decreases of 10% to 49% in the V99 occurred in multiple patients for whomthe beampaths traversed the lung-diaphragminterface and were, therefore,more sensitive to respiration-induced changes in the water equivalent path length. Fractionation alone apparently did not mitigate the interplay effect beyond 6 fractions. Conclusion: The interplay effect is not a concern when delivering scanning proton beams to younger pediatric patients with tumors located in the retroperitoneal space and tumor motion of ,5 mm. Children and adolescents with diaphragmatic tumor motion exceeding 10 mm require special attention, because significant declines in target coverage and dose homogeneity were seen in simulated treatments of such patients.

Cite

CITATION STYLE

APA

Boria, A. J., Uh, J., Pirlepesov, F., Stuckey, J. C., Axente, M., Gargone, M. A., & Hua, C. H. (2018). Interplay effect of target motion and pencil-beam scanning in proton therapy for pediatric patients. International Journal of Particle Therapy, 5(2), 1–10. https://doi.org/10.14338/IJPT-17-00030.1

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free