Abstract
Background: Investigate the effectiveness of beam-hardening (BH) correction using ECG-gated dual-kVp computed tomography (CT) for myocardial imaging. Methods and Results: A human heart specimen and artificial descending aorta were scanned using both dual-kVp and single 120-kVp modes. The myocardial CT values at the anterior and posterobasal walls were 89.4±6.2 and 87.1±6.9 HU (P=0.59) for the 69-keV images, and 88.3±5.9 and 49.9±13.5 HU (P=0.002) for the 120-kVp images, respectively. Conclusions: Fast-switching dual-kVp technology is feasible for the correction of CT value deficits induced by BH that mimics perfusion defects.
Author supplied keywords
Cite
CITATION STYLE
Yamada, M., Jinzaki, M., Kuribayashi, S., Imanishi, N., Funato, K., & Aiso, S. (2012). Beam-hardening correction for virtual monochromatic imaging of myocardial perfusion via fast-switching dual-kVp 64-slice computed tomography: - A pilot study using a human heart specimen -. Circulation Journal, 76(7), 1799–1801. https://doi.org/10.1253/circj.CJ-12-0463