Abstract
Background:A T-piece resuscitator (TPR) uses a built-in manometer to set the inflation pressures, but we are not informed what pressures are actually delivered distally. Aim of this study was to measure the proximal and distal pressures under different gas conditions when using a TPR.Methodology/Findings:A test lung was ventilated using a TPR (PIP 25 cmH2O, PEEP 5 cmH2O) with a gas flow rate of 8 L/min. A) Pressure delivered by six different TPRs was tested. To test variability 20 participants were asked to set PEEP and PIP pressures to 25/5 cmH2O. B) PIP and PEEP were measured proximal and distal of the TPR when using standard tubing or heated tubing with or without a humidifier. In experiment A mean (SD) proximal PIP and PEEP of the TPRs were respectively 20.3 (0.3) cmH2O (19.9-20.6 cmH2O) and 4.9 (0.1) cmH2O. When 20 participants set pressures; PIP 26.7 (0.5) cm H2O and PEEP 5.9 (0.44) cmH2O were measured. Experiment B showed that the decrease of PIP between proximal and distal pressures was not clinically significant. However there was a significant decrease of PEEP using the standard tubing (5.1 (0.1) cmH2O proximally versus 4.8 (0.2) cmH2O distally; p<0.001) compared to, when using a humidifier with associated tubing and the humidifier turned on, 5.1 (0.1) proximally versus 3.9 (0.2) cmH2O distally; (p<0.001).Conclusion/Significance:The accuracy of the built-in manometer of a TPR is acceptable. Most pressures set proximally are comparable to the actual pressures delivered distally. However, when using tubing associated with the humidifier PEEP decreases distally by 1.1-1.2 cmH2O and users should anticipate on this. © 2013 Roegholt et al.
Cite
CITATION STYLE
Roegholt, E., van Vonderen, J. J., Walther, F. J., Roehr, C. C., & te Pas, A. B. (2013). Do We Deliver the Pressures We Intend to When Using a T-Piece Resuscitator? PLoS ONE, 8(5). https://doi.org/10.1371/journal.pone.0064706
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.