Quadrupolar ion excitation for radiofrequency-only mass filter operation

5Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Trajectory calculations are used to model a mass filter based on the radiofrequency (rf)-only operation of a linear quadrupole with resonantquadrupole excitation of ions (resonant excitation applied with the same spatial electric field as the main quadrupole rf field). Ions are not trapped, but pass continuously through the quadrupole. Excited ions gain axial kinetic energy in the fringe field at the quadrupoleexit, overcome a stopping potential and are transmitted to an external detector. No quadrupole direct current is required, unlike conventional operation at the tip of the first stability diagram. Quadrupole excitation can be applied with amplitude or frequencymodulation of the main rf voltage, or with an auxiliary excitation voltage. All three methods give the same mass resolution. The mass resolution, R, is given by R≈ 0.5q(dβ/dq)n where q is a Mathieu parameter, β (q), determines the frequency of ion oscillation and n is the number of cycles of the rf field experienced by an ion, determined by the flight time through the quadrupole. A disadvantage of this mode of operation is that the flight times of the ions and the excitation amplitudes or modulation depths need to be synchronized. © 2014 IM Publications LLP.

Cite

CITATION STYLE

APA

Douglas, D. J., Polyakov, A., & Konenkov, N. V. (2014). Quadrupolar ion excitation for radiofrequency-only mass filter operation. European Journal of Mass Spectrometry, 20(3), 207–214. https://doi.org/10.1255/ejms.1278

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free