Impact of movement and motion-artefact correction on image quality and interpretability in CBCT units with aligned and lateral-offset detectors

14Citations
Citations of this article
32Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Objectives: To evaluate the impact of movement and motion-artefact correction systems on CBCT image quality and interpretability of simulated diagnostic tasks for aligned and lateral-offset detectors. Methods: A human skull simulating three diagnostic tasks (implant planning in the anterior maxilla, implant planning in the left-side- mandible and mandibular molar furcation assessment in the right-side- mandible) was mounted on a robot performing six movement types. Four CBCT units were used: Cranex 3Dx (CRA), Ortophos SL (ORT), Promax 3D Mid (PRO), and X1. Protocols were tested with aligned (CRA, ORT, PRO, and X1) and lateral-offset (CRA and PRO) detectors and two motion-artefact correction systems (PRO and X1). Movements were performed at one moment-in- time (t1), for units with an aligned detector, and three moments-in- time (t1-first- half of the acquisition, t2-second- half, t3-both) for the units with a lateral-offset detector. 98 volumes were acquired. Images were scored by three observers, blinded to the unit and presence of movement, for motion-related stripe artefacts, overall unsharpness, and interpretability. Fleiss' κ was used to assess interobserver agreement. Results: Interobserver agreement was substantial for all parameters (0.66-0.68). For aligned detectors, in all diagnostic tasks a motion-artefact correction system influenced image interpretability. For lateral-offset detectors, the interpretability varied according to the unit and moment-in-time, in which the movement was performed. PRO motion-artefact correction system was less effective for the offset detector than its aligned counterpart. Conclusion: Motion-artefact correction systems enhanced image quality and interpretability for units with aligned detectors but were less effective for those with lateral-offset detectors.

Cite

CITATION STYLE

APA

Santaella, G. M. H., Wenzel, A., Haiter-Neto, F., Rosalen, P. L., & Spin-Neto, R. (2019). Impact of movement and motion-artefact correction on image quality and interpretability in CBCT units with aligned and lateral-offset detectors. Dentomaxillofacial Radiology, 49(1). https://doi.org/10.1259/dmfr.20190240

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free