Precision and agreement of corneal power measurements obtained using a new corneal topographer ophthatop

12Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

Abstract

Purpose: To evaluate repeatability and reproducibility of anterior corneal power measurements obtained with a new corneal topographer OphthaTOP (Hummel AG, Germany) and agreement with measurements by a rotating Scheimpflug camera (Pentacam HR, Oculus, Germany) and an automated keratometer (IOLMaster, Carl Zeiss Meditec, Germany). Methods: The right eyes of 79 healthy subjects were prospectively measured three times with all three devices. Another examiner performed three additional scans with the OphthaTOP in the same session. Within one week, the first examiner repeated the measurements using the OphthaTOP. The flat simulated keratometry (Kf), steep K (Ks), mean K (Km), J0, and J45 were noted. Repeatability and reproducibility of measurements were assessed by within-subject standard deviation (Sw), repeatability (2.77 Sw), coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement between devices was assessed using 95% limits of agreement (LoA). Results: Intraobserver repeatability and interobserver and intersession reproducibility of all measured parameters showed a 2.77 Sw of 0.29 diopter or less, a CoV of less than 0.24%, and an ICC of more than 0.906. Statistically significant differences (P<0.001) were found between the parameters analyzed by the three devices, except J0 and J45. The mean differences between OphthaTOP and the other two devices were small, and the 95% LoA was narrow for all results. Conclusions: The OphthaTOP showed excellent intraobserver repeatability and interobserver and intersession reproducibility of corneal power measurements. Good agreements with the other two devices in these parameters were found in healthy eyes.

Cite

CITATION STYLE

APA

Huang, J., Savini, G., Chen, H., Bao, F., Li, Y., Chen, H., … Wang, Q. (2015). Precision and agreement of corneal power measurements obtained using a new corneal topographer ophthatop. PLoS ONE, 10(1). https://doi.org/10.1371/journal.pone.0109414

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free