Abstract

An optical cavity-based sensor using a differential detection method has been proposed for point-of-care diagnostics. We developed a low-cost and portable optical cavity-based sensor system using a 3D printer and ff-the-shelf optical components. In this paper, we demonstrate the sensing capability of the portable system through refractive index measurements. Fabricated optical cavity samples were tested using the portable system and compared to simulation results. A referencing technique and digital low pass filtering were applied to reduce the noise of the portable system. The measurement results match the simulation results well and show the improved linearity and sensitivity by employing the differential detection method. The limit of detection achieved was 1.73 X 10-5 Refractive Index Unit (RIU), which is comparable to other methods for refractive index sensing.

Cite

CITATION STYLE

APA

Rho, D., Breaux, C., & Kim, S. (2019). Demonstration of a low-cost and portable optical cavity-based sensor through refractive index measurements. Sensors (Switzerland), 19(9). https://doi.org/10.3390/s19092193