Comparison of DMA-80 and ICP-MS Combined with Closed-Vessel Microwave Digestion for the Determination of Mercury in Coal

23Citations
Citations of this article
36Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

As one of the most widely used techniques for concentration determination of trace elements in coal, inductively coupled plasma mass spectrometry (ICP-MS) has also been used in several studies for the determination of mercury concentration in coal. ICP-MS after closed-vessel microwave digestion and a Milestone DMA-80 are employed in this study to determine the mercury concentration in coal. Three NIST standard references of coal samples were selected as references to verify the accuracy of the test results. The Au rinse solution (200 μg/L, 5% HNO3) can diminish mercury memory effects to a blank level within 80 seconds. The results showed that ICP-MS can accurately determine the mercury content in mercury standard solutions, but the mercury concentration in most NIST samples after microwave digestion is lower than the detection level of the ICP-MS. The inaccuracy may be due to volatilization of mercury during solid sample digestion process. By contrast, the determined concentrations in NIST samples by the Milestone DMA-80 are very close to the verified values. Therefore, ICP-MS is not recommended to analyze mercury in coal after digestion even in a closed-vessel digestion system, but the mercury direct analyzer (without digestion) is recommended to analyze mercury in coal.

Cite

CITATION STYLE

APA

Zhang, S., & Zhou, M. (2020). Comparison of DMA-80 and ICP-MS Combined with Closed-Vessel Microwave Digestion for the Determination of Mercury in Coal. Journal of Analytical Methods in Chemistry, 2020. https://doi.org/10.1155/2020/8867653

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free