Abstract
The COVID-19 pandemic has highlighted resource constraints in respiratory support. The oxygen transfer characteristics of a specific hollow fiber membrane dialyser was investigated with a view to repurposing the device as a low-cost, readily available blood oxygenator. Oxygen transfer in a low-flux hollow fiber dialyser with a polysulfone membrane was studied by passing first water and then blood through the dialyser in countercurrent to high-purity oxygen. Oxygen transfer rates of about 15% of the nominal 250 ml (STP)/min of a typical adult oxygen consumption rate were achieved for blood flow rates of 500 ml/min. Using two such dialysis devices in parallel could provide up to 30% of the nominal oxygen consumption. Specific hollow fiber dialysis devices operating with suitable pumps in a veno-venous access configuration could provide a cost-effective and readily available supplementation of respiratory support in the face of severe resource constraints.
Cite
CITATION STYLE
Rubin, D. M., Stacey, N. T., Matambo, T., Vale, C. D., Sussman, M. J., Snyman, T., … Hildebrandt, D. (2020). Toward Respiratory Support of Critically Ill COVID-19 Patients Using Repurposed Kidney Hollow Fiber Membrane Dialysers to Oxygenate the Blood. Journal of Healthcare Engineering, 2020. https://doi.org/10.1155/2020/8862645