Comparison of a micro-electro-mechanical system airflow sensor with the pneumotach in the forced oscillation technique

2Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

Abstract

This study supports the use of thin-film micro-electro-mechanical system (MEMS) airflow sensors in the forced oscillation technique. Materials and methods: The study employed static testing using air flow standards and computer-controlled sound attenuations at 8 Hz. Human feasibility studies were conducted with a testing apparatus consisting of a pneumotach and thin-film MEMS air flow sensors in series. Short-time Fourier transform spectra were obtained using SIGVIEW software. Results: Three tests were performed, and excellent correlations were observed between the probes. The thin-film MEMS probe showed superior sensitivity to higher frequencies up to 200 Hz. Conclusion: The results suggest that lower-cost thin-film MEMS can be used for forced oscillation technique applications (including home care devices) that will benefit patients suffering from pulmonary diseases such as asthma, COPD, and cystic fibrosis.

Cite

CITATION STYLE

APA

Xu, X. K., Harvey, B. P., Lutchen, K. R., Gelbman, B. D., Monfre, S. L., Coifman, R. E., & Forbes, C. E. (2018). Comparison of a micro-electro-mechanical system airflow sensor with the pneumotach in the forced oscillation technique. Medical Devices: Evidence and Research, 11, 419–426. https://doi.org/10.2147/MDER.S181258