Pacing and sensing optimization of permanent His-bundle pacing in cardiac resynchronization therapy/implantable cardioverter defibrillators patients: Value of integrated bipolar configuration

46Citations
Citations of this article
51Readers
Mendeley users who have this article in their library.

Abstract

Aims The clinical implementation of His-bundle pacing (HBP) has been limited by concern over higher capture thresholds (CTs) and lower R-wave amplitudes (RWAs), when compared with right ventricular (RV) pacing. The aim of this study was to assess the optimal pacing configuration for HBP lead when incorporated with cardiac resynchronization therapy/ implantable cardioverter defibrillators (CRT-D/ICD) by testing of an integrated bipolar configuration. Methods and results HBP was achieved in 25 CRT-D and 13 ICD patients at implantation. Their RWA, CT@0.5 ms and impedance with His-bundle (HB) tip-RV coil, and HB unipolar and bipolar (tip-ring) configurations were measured acutely. Capture threshold in CRT-D patients was that needed to 'correct' complete left bundle block. Among all 16 CRT-D patients with permanent HBP, theirRWA, CT@0.5 ms, and pacing impedance with HBP tip-RV coil and HBP tip-ring were measured at 1 month and 3 months follow-up. HigherRWA and lower CT were achieved with HB tip-RV coil configuration than either of the other two configurations at implantation (n = 38, P < 0.05, respectively). At 1 and 3 months followup for patients receiving permanent HBP, RWA and CT with HB tip-RV coil configuration remained stable and better than those with HB tip-ring configuration (n = 16, P < 0.05, respectively). Furthermore, HB lead pulse energy delivered with HB tip-RV coil configuration was also less than with HB tip-ring at 3 months (n = 16, P = 0.017). Conclusion Incorporation of HBP into a CRT-D/ICD system is feasible, and pacing thresholds/sensing can be optimized using a novel integrated bipolar configuration with the RV lead.

Cite

CITATION STYLE

APA

Su, L., Xu, L., Wu, S. J., & Huang, W. J. (2016). Pacing and sensing optimization of permanent His-bundle pacing in cardiac resynchronization therapy/implantable cardioverter defibrillators patients: Value of integrated bipolar configuration. Europace, 18(9), 1399–1405. https://doi.org/10.1093/europace/euv306

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free